Guide
Bias at clinical decision limits: what a method comparison misses
When a laboratory changes analyser, the method comparison asks whether the two agree. The clinician asks something else: whose result now lands on the other side of a decision limit. This guide answers both.
Two questions, not one
A method comparison answers a numerical question: how far apart are results from the old and the new measurement procedure, and is that within what the laboratory will accept [CLSI EP09]. Near a decision limit there is a second, interpretive question: does the same patient land in the same category on both? The two can disagree. Numerically similar results can fall on opposite sides of a limit, and a difference outside the allowable error can leave a patient’s category unchanged [Cai et al., 2026].
Clinical decisions are made at limits, not at averages. A percent difference answers the first question. Clinicians and quality managers need the second.
The comparison line: Passing-Bablok and Bland-Altman in brief
Passing-Bablok regression fits a straight line of new results on old without assuming either method is free of error, and is robust to outliers [Passing and Bablok, 1983]. Its slope says whether the difference grows with concentration (proportional) and its intercept whether there is a fixed offset (constant). Both come with intervals, usually from the method’s own procedure or from bootstrap resampling.
Bland-Altman analysis plots each pair’s difference against its mean and reports the average difference with limits of agreement, the range in which most differences fall [Bland and Altman, 1986]. It summarises agreement across the whole range.
Both are standard tools for verifying comparability in a medical laboratory, alongside Deming regression and paired difference tests [Abdel Ghafar and El-Masry, 2021]. Neither tells you what happens at a particular limit.
Bias at the limit, with an interval
With a comparison line in hand, the systematic error at any concentration can be read off it: that is how method-validation practice estimates error at medical decision concentrations rather than as one average [Westgard]. At a decision limit L, with slope b and intercept a:
bias at L = (b × L + a) − L
Expressed as a percentage of L, this is the figure to set against the laboratory’s acceptance limit at that concentration. Repeating the fit on bootstrap resamples of the pairs gives an interval for it. Report bias at each limit the laboratory publishes, not once for the whole range. A proportional difference that looks small on average can be large at a high limit, and a constant offset matters most at a low one.
Counting patients across a line
Bias at a limit describes the method. The clinician wants a count: how many patients were in one category on the old analyser and in another on the new. Place each pair’s old and new result into the categories the limits define, and count the pairs whose category differs, separately for each limit and each direction.
Direction matters. A new analyser that reads lower moves patients down across a limit; one that reads higher moves them up; a change in slope can do both at different limits. A table of old category by new category shows all of it at once.
Why repeat testing sets the baseline
Some of those crossings happen with no change at all. Measure a sample twice on one analyser and a result close to a limit can land on either side of it, purely from imprecision. A raw count of crossings overstates what the change did.
The baseline comes from the analyser’s repeat-test imprecision (its coefficient of variation). For each result, take the probability that a repeat measurement on the same analyser would put it in a different category; summed over all results, that is the number of crossings to expect from repeat testing alone. The excess of observed crossings over that expectation estimates the change’s own effect; whether it is more than chance takes an exact test at a measured imprecision, and at a default imprecision the excess only describes the counts. Measure the imprecision from QC or duplicate data where the laboratory has it, and state it when a default is used.
This gives a re-test band. Results that land within a narrow range of a limit on the new analyser are the ones most likely to have crossed, by change or by chance. Re-testing them during the transition catches much of the excess for a known share of extra tests, and that share can be fixed in advance as a budget.
A worked example
Sample
The example uses a public CD4 method comparison: 1,885 paired counts from an open-access deposit in which one flow-cytometry platform replaced another [Coetzee and Glencross, 2017]. We ran it as a laboratory export, with decision limits at 100, 200, 350, 500 cells/µL and a default repeat-test imprecision.
Comparison line with slope 0.867 (95% interval 0.857 to 0.878) and intercept -1.2, drawn against the line of no change. Decision limits drawn at 100, 200, 350, 500 cells/µL. Where the comparison line sits above the line of no change, the new system reads higher; below, lower.
Fig. 1. Comparison line for the CD4 sample: slope 0.87, so the new platform reads about 14% lower on average, with the difference growing with the count.
Source Coetzee LM, Glencross DK (2017). PLoS One. doi:10.1371/journal.pone.0187456 (opens in a new tab); PMC5669480 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Re-analysed by NakedSignal.
Show the numbers
| Pairs | 1,885 |
|---|---|
| Slope with its interval | 0.867 (0.857 to 0.878) |
| Intercept | -1.2 |
| Decision limits | 100, 200, 350, 500 cells/µL |
All decision limits combined
Crossed between the two systems: 535
Expected from repeat testing alone: about 238
Fig. 2. Across the decision limits, 535 results changed category against about 238 expected from repeat testing alone.
Source Same deposit, CC BY 4.0.
Without adjustment, the change moves 535 results across a limit where repeat testing alone would move about 238. Correcting with the comparison line and re-testing results inside a band around each limit removed 66% of that excess on pairs the band was not fitted to, against 46% for the correction line alone, while re-testing 15% of results. The sample report shows the bias at each limit read off the comparison line, and the re-test band.
Related checks
- Commutability. Whether control or EQA material behaves like patient samples on both methods is a separate question; this method works on patient pairs.
- Reagent and calibrator lot changes. Lot studies are usually too small for this count: it needs at least 20 pairs, and in our simulations at 20 to 40 pairs it rarely flags a 5% lot shift that a paired t-test on the same pairs finds, so a single lot is judged on the mean difference at each limit. See the guide to lot changes.
- ISO 15189 verification. A count of patients across limits can be added to a laboratory’s verification record and to its note to clinicians.
Sources
- Cai X, Yang W, Lu Q, Lin Y (2026). Numerical and interpretive comparability of optical and mechanical coagulation analyzers: a Zlog-supported method comparison study. Diagnostics (Basel). pmc.ncbi.nlm.nih.gov/articles/PMC13511516/
- Clinical and Laboratory Standards Institute. EP09: Measurement procedure comparison and bias estimation using patient samples. CLSI guideline, third edition. clsi.org/standards/products/method-evaluation/documents/ep09/
- Passing H, Bablok W (1983). A new biometrical procedure for testing the equality of measurements from two different analytical methods. Part I. Journal of Clinical Chemistry and Clinical Biochemistry. doi.org/10.1515/cclm.1983.21.11.709
- Bland JM, Altman DG (1986). Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. doi.org/10.1016/s0140-6736(86)90837-8
- Abdel Ghafar MT, El-Masry MI (2021). Verification of quantitative analytical methods in medical laboratories. Journal of Medical Biochemistry. pmc.ncbi.nlm.nih.gov/articles/PMC8199534/
- Westgard JO. The comparison of methods experiment. Westgard QC, Basic Method Validation. www.westgard.com/lessons/basic-method-validation/lesson23.html
- Coetzee LM, Glencross DK (2017). PLoS One. doi.org/10.1371/journal.pone.0187456