NakedSignal

Guide

Jaffe or enzymatic creatinine: what a method change does at the decision lines

Creatinine is measured by two chemistries that agree on average and differ in what else they react to. This guide covers why, what calibration traceability fixes, and how to count the results a change of method or device moves across a decision line.

Why the two chemistries differ

The Jaffe method measures the colour creatinine forms with alkaline picrate. Other substances in the sample react too, so the method is less specific. Enzymatic methods use a chain of enzyme reactions and are less prone to those interferences. In a comparison of both on the same analyser and samples, the authors describe the Jaffe method as less expensive but more susceptible to interferences [Schmidt et al., 2015, PLoS ONE]. The KDIGO 2024 guideline reinforces the use of enzymatic assays instead of the Jaffe method, given the interference of many drugs and substances with the latter, and asks laboratories to consider enzymatic assays consistently in children [KDIGO, 2024].

What IDMS traceability fixes

Creatinine calibration is standardised by making routine methods traceable to a reference measurement procedure; the laboratory Jaffe method in one of the comparisons below is traceable to isotope dilution mass spectrometry (IDMS) [Msilanga et al., 2025]. The US National Kidney Disease Education Program’s laboratory working group set out a plan for standardising creatinine measurement worldwide, and noted that measurement variability makes estimated GFR least accurate in the normal and slightly raised creatinine range, which is the range that matters for detecting chronic kidney disease [Myers et al., 2006].

Calibration is the part it fixes. Analytical interferences remain [Myers et al., 2006]. Two methods traceable to the same reference can agree on a clean sample and still differ on a patient whose sample carries an interfering substance, so a method change is checked on patient samples as well as calibrators.

How a small bias moves eGFR categories

eGFR is calculated from creatinine with age and sex, and rises as creatinine falls. A small creatinine bias therefore moves eGFR most where results are dense around a category boundary. In the outpatient comparison of the two methods, there was no overall mean bias, yet a small share of pairs fell into different eGFR categories, with the highest risk at the boundary that defines chronic kidney disease; the differences were not significant relative to biological variation, and the authors judged the risk low if results near that boundary are interpreted with caution [Schmidt et al., 2015, PLoS ONE].

That is the right shape of answer for a method change: how many results move across a line, and how that compares with what repeat testing alone would move at the laboratory’s measured imprecision. The guide on bias at clinical decision limits sets out the counting method.

Five public comparisons

Public data

The explorer holds five paired creatinine comparisons from four open-access studies: one comparison of two chemistries run on the same samples on one laboratory analyser, read as a method change, and four point-of-care devices against a laboratory. We re-analysed each the same way, at the same two creatinine lines. The expected count for the method change uses the imprecision its study measured; the four point-of-care comparisons use a default repeat-test imprecision.

Public data

Same analyser, kinetic Jaffe creatinine against central lab analyser, enzymatic creatinine (Schmidt et al., 2015, PLoS ONE)

Crossed between the two systems: 23

Expected from repeat testing alone: about 12

Point of care creatinine meter against central lab analyser, Jaffe creatinine (Msilanga et al., 2025)

Crossed between the two systems: 29

Expected from repeat testing alone: about 19

Point of care device B against central lab chemistry analyser (Giachino et al., 2024)

Crossed between the two systems: 46

Expected from repeat testing alone: about 16

Point of care analyser, venous whole blood against central lab serum analyser, kinetic Jaffe (Krisher et al., 2024)

Crossed between the two systems: 1

Expected from repeat testing alone: about 1

Point of care creatinine meter, capillary against central lab serum analyser, kinetic Jaffe (Krisher et al., 2024)

Crossed between the two systems: 28

Expected from repeat testing alone: about 1

Fig. 1. For the two chemistries on one analyser, read as a switch from enzymatic to Jaffe, 23 results changed category against about 12 expected from repeat testing alone. That expected count uses the 1.8% imprecision the study itself measured; at that imprecision the count from repeat testing alone falls between 6 and 17 with 95% probability, so the 23 crossings are about twice the expected count, and by the exact Poisson-binomial test 23 or more from repeat testing alone has probability 0.00017.That 1.8% is also our program's laboratory-analyser default (1.79%, this study's own reported precision), so the default gives about the same count; the program's cautious point-of-care default of 6%, which it uses when a point-of-care device is involved and no precision data are supplied, would give about 41, which overstates repeat-testing noise for these laboratory analysers. At the upper end of the study's own precision estimate, 2.34%, about 15 would be expected and 23 or more has probability 0.014. Of the rest, 3 of the 4 point-of-care comparisons sat above the repeat-testing count at a default imprecision, which describes the counts and is not a test.

Source Schmidt RL, Straseski JA, Raphael KL, Adams AH, Lehman CM (2015). A Risk Assessment of the Jaffe vs Enzymatic Method for Creatinine Measurement in an Outpatient Population. PLoS One. doi:10.1371/journal.pone.0143205 (opens in a new tab); PMC4657986 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Msilanga D, Muiru A, Msangi E, Shoo J, Mngumi J, Komba E, Balandya E, Ruggajo P, Bhimma R, Liu K (2025). PLoS One. doi:10.1371/journal.pone.0331969 (opens in a new tab); PMC12425315 (opens in a new tab). Licence: CC0 1.0 (opens in a new tab). Giachino M, Vetter B, Perone SA, Correia JC, Erkosar B, Heller O, Khanal VK, Lab B, Pataky Z, Poudel S, Rai M, Sharma SK (2024). Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study. PLOS Glob Public Health. doi:10.1371/journal.pgph.0003760 (opens in a new tab); PMC11449279 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Krisher L, Jaramillo D, Dye-Robinson A, Dally M, Butler-Dawson J, Brindley S, Pilloni D, Cruz A, Villarreal Hernandez K, Schaeffer J, Adgate JL, Newman LS (2024). Application and comparison of point-of-care devices for field evaluation of underlying health status of Guatemalan sugarcane workers. PLOS Glob Public Health. doi:10.1371/journal.pgph.0003380 (opens in a new tab); PMC11265697 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Re-analysed by NakedSignal.

Show the numbers
ComparisonPairsSlopeLinesChanged categoryExpected from repeat testingRe-test band
Same analyser, kinetic Jaffe creatinine against central lab analyser, enzymatic creatinine (Schmidt et al., 2015, PLoS ONE)5290.94 (95% interval 0.91 to 0.97)1.2 and 1.5 mg/dL23about 12Full
Point of care creatinine meter against central lab analyser, Jaffe creatinine (Msilanga et al., 2025)3581.08 (95% interval 1.03 to 1.14)106.08 and 132.6 µmol/L29about 19Full
Point of care device B against central lab chemistry analyser (Giachino et al., 2024)3451.11 (95% interval 0.97 to 1.21)1.2 and 1.5 mg/dL46about 16Full
Point of care analyser, venous whole blood against central lab serum analyser, kinetic Jaffe (Krisher et al., 2024)891.43 (95% interval 1.25 to 1.67)1.2 and 1.5 mg/dL1about 1Indicative
Point of care creatinine meter, capillary against central lab serum analyser, kinetic Jaffe (Krisher et al., 2024)894.33 (95% interval 3.12 to 6.41)1.2 and 1.5 mg/dL28about 1Indicative

Two chemistries on one analyser, read as a method change, at the study’s own precision

In the outpatient study, both chemistries ran on the same samples on the same analyser [Schmidt et al., 2015, PLoS ONE]. Our run treats the enzymatic method as the current one and the Jaffe method as the new one. Across 529 pairs, 23 results changed category, against about 12 expected from repeat testing alone. That expected count uses the 1.8% imprecision the study itself measured; at that imprecision the count from repeat testing alone falls between 6 and 17 with 95% probability, so the 23 crossings are about twice the expected count, and by the exact Poisson-binomial test 23 or more from repeat testing alone has probability 0.00017. That 1.8% is also our program's laboratory-analyser default (1.79%, this study's own reported precision), so the default gives about the same count; the program's cautious point-of-care default of 6%, which it uses when a point-of-care device is involved and no precision data are supplied, would give about 41, which overstates repeat-testing noise for these laboratory analysers. At the upper end of the study's own precision estimate, 2.34%, about 15 would be expected and 23 or more has probability 0.014. Separating the crossings seen at a decision limit into the part imprecision alone would give and the part due to the change is a published procedure (Schmidt et al., 2015, Clin Chim Acta), which the creatinine study we use applies. What we add is a program that runs it on a laboratory’s own comparison and precision data at its own decision limits, keeps the re-test list in the laboratory and releases files designed to carry only aggregates. The comparison line (Fig. 2) has a slope of 0.94 (95% interval 0.91 to 0.97), so the difference is small at the lines but not zero. The paper reports no overall mean difference in creatinine; a Deming or least-squares fit to the same pairs gives a slope close to one, and the lower Passing-Bablok slope reflects curvature at low creatinine.

Public data
Comparison line, creatinine (mg/dL)Comparison line with slope 0.939 (95% interval 0.914 to 0.967) and intercept 0.053, drawn against the line of no change. Decision limits drawn at 1.2, 1.5 mg/dL. Where the comparison line sits above the line of no change, the new system reads higher; below, lower.11.511.51.21.5No changeComparison lineOld system, mg/dLNew system, mg/dL

Comparison line with slope 0.939 (95% interval 0.914 to 0.967) and intercept 0.053, drawn against the line of no change. Decision limits drawn at 1.2, 1.5 mg/dL. Where the comparison line sits above the line of no change, the new system reads higher; below, lower.

Fig. 2. Enzymatic and Jaffe creatinine on the same analyser: slope 0.94, close to the line of no change across the range of the two creatinine lines.

Source Schmidt RL, Straseski JA, Raphael KL, Adams AH, Lehman CM (2015). A Risk Assessment of the Jaffe vs Enzymatic Method for Creatinine Measurement in an Outpatient Population. PLoS One. doi:10.1371/journal.pone.0143205 (opens in a new tab); PMC4657986 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Re-analysed by NakedSignal.

Point of care against the laboratory: more to manage

The point-of-care comparisons are a different kind of change: a device outside the laboratory, often on capillary blood, against a laboratory method. KDIGO supports point-of-care creatinine where laboratory access is limited, provided the same quality criteria, including external quality assessment, apply [KDIGO, 2024].

  • A point-of-care meter against a laboratory Jaffe method in people living with HIV: the authors report substantial agreement and a small eGFR bias that changes direction across the range [Msilanga et al., 2025]. Here 29 results changed category against about 19 expected (default CV), at 106.08 and 132.6 µmol/L (the same two lines in SI units).
  • A point-of-care device against a central laboratory analyser: the authors report a small negative average bias within recommended limits [Giachino et al., 2024]. Here 46 results changed category against about 16 expected (default CV).
  • Two devices against a laboratory kinetic Jaffe method in field conditions: the authors found the venous analyser agreed well and the capillary meter read higher, especially at higher values [Krisher et al., 2024]. The venous analyser moved 1 result against about 1 expected (default CV); the capillary meter moved 28 against about 1 (default CV). Both comparisons are small, so their re-test bands are graded indicative.

Each comparison has its own page with its comparison line, re-test band and full numbers (links in the table under Fig. 1), and all of them sit side by side in the change explorer.

Choosing the lines

The creatinine lines in these comparisons (1.2 and 1.5 mg/dL) are set once across the public set so every series reads the same way. Clinical guidelines work in eGFR categories, which need age and sex with each result. A laboratory running its own comparison names the lines its clinicians use, for creatinine, eGFR or both, before the results are opened.

  • Paediatrics. Children have low creatinine, where non-creatinine chromogens weigh most in the Jaffe method [KDIGO, 2024]. A paediatric service needs a paediatric comparison; none of the public series here is one.
  • Choice of eGFR equation. The equation moves eGFR categories independently of the creatinine method, so an equation change deserves its own count.
  • Specific interferents. Which drugs or substances affect which method is a question for interference studies; paired patient comparisons show their combined effect.

Sources

  1. Schmidt RL, Straseski JA, Raphael KL, Adams AH, Lehman CM (2015). A Risk Assessment of the Jaffe vs Enzymatic Method for Creatinine Measurement in an Outpatient Population. PLoS ONE. pmc.ncbi.nlm.nih.gov/articles/PMC4657986/
  2. Myers GL, Miller WG, Coresh J, et al. (2006). Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. Clinical Chemistry. doi.org/10.1373/clinchem.2005.0525144
  3. Levin A, Ahmed SB, Carrero JJ, et al. (2024). Executive summary of the KDIGO 2024 Clinical Practice Guideline for the Evaluation and Management of Chronic Kidney Disease: known knowns and known unknowns. Kidney International. kdigo.org/wp-content/uploads/2017/02/KDIGO-2024-CKD-Guideline-Executive-Summary.pdf
  4. Giachino M, Vetter B, Perone SA, et al. (2024). Performance and usability of cardiometabolic point of care devices in Nepal: A prospective, quantitative, accuracy study. PLOS Global Public Health. pmc.ncbi.nlm.nih.gov/articles/PMC11449279/
  5. Msilanga D, Muiru A, Msangi E, et al. (2025). PLoS ONE. pmc.ncbi.nlm.nih.gov/articles/PMC12425315/
  6. Krisher L, Jaramillo D, Dye-Robinson A, et al. (2024). Application and comparison of point-of-care devices for field evaluation of underlying health status of Guatemalan sugarcane workers. PLOS Global Public Health. pmc.ncbi.nlm.nih.gov/articles/PMC11265697/