Point-of-care creatinine in venous blood vs a Jaffe laboratory analyser

Public dataA public method-comparison deposit, not a laboratory's data. Systems are described generically; the source names them.

Central lab serum analyser, kinetic Jaffe replaced by point of care analyser, venous whole blood: 89 paired results from a published study by Krisher et al. (2024), re-analysed by NakedSignal as if the study were one laboratory. At 1.2, 1.5 mg/dL, 1 results changed category after the change, against about 1 expected from repeat testing alone. Here the change moved fewer results than repeat testing would, and the report says so.

run 25 September 2026 · config 6eb23df45893 · run file sha256 5586cd892ef1

Chemistry

Creatinine, mg/dL

Central lab serum analyser, kinetic Jaffe replaced by Point of care analyser, venous whole blood. 89 paired results.

Compare it with the other comparisons in the explorer

Comparison line

The new system reads 0.6% lower than the old system on average. The comparison line has a slope of 1.429 (95% interval 1.25 to 1.667) and an intercept of -0.343.

Public data
Comparison line, creatinine (mg/dL)Comparison line with slope 1.429 (95% interval 1.25 to 1.667) and intercept -0.343, drawn against the line of no change. Decision limits drawn at 1.2, 1.5 mg/dL. Where the comparison line sits above the line of no change, the new system reads higher; below, lower.0.511.5211.51.21.5No changeComparison lineOld system, mg/dLNew system, mg/dL

Fig. 1. The comparison line against the line of no change, over the range of the decision limits (1.2, 1.5 mg/dL). Dots mark where each line meets each limit; the gap between them is the bias at that limit.

run 25 September 2026 · config 6eb23df45893 · run file sha256 5586cd892ef1

Crossing each limit

Results that moved across a decision limit when the new system replaced the old, against the number repeat testing on one unchanged system would move. The program reports the count across all limits together.

All decision limits together (1.2, 1.5 mg/dL)

Crossed after the change: 1

Expected from repeat testing alone: about 1

Grade and warnings

  • Re-test band: Indicative: 89 pairs, below the 100-pair protocol minimum.
  • Band half-width around each limit: 27.66%.
  • Results re-tested on data the band was not fitted to, median: 13%.
  • Repeat-test imprecision: assumed, not measured at this site.
  • Warning from the program: The change moved fewer than 5 more people across a line than repeat testing alone would; excess caught is not meaningful here.

Source

Krisher L, Jaramillo D, Dye-Robinson A, Dally M, Butler-Dawson J, Brindley S, Pilloni D, Cruz A, Villarreal Hernandez K, Schaeffer J, Adgate JL, Newman LS (2024). Application and comparison of point-of-care devices for field evaluation of underlying health status of Guatemalan sugarcane workers. PLOS Glob Public Health. doi:10.1371/journal.pgph.0003380 (opens in a new tab); PMC11265697 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Changes: Paired results re-analysed by NakedSignal's program; instruments described generically; only derived results are shown. The deposit names the systems.

Show the numbers
All values for creatinine, Central lab serum analyser, kinetic Jaffe to Point of care analyser, venous whole blood
FieldValue
Paired results used89
Comparison line slope1.428571
Slope, 95% interval1.25 to 1.666667
Comparison line intercept-0.342857
Mean difference, new minus old, as % of old-0.556
Decision limits (mg/dL)1.2, 1.5
Crossed after the change1
Expected from repeat testing alone1.25
Band half-width (%)27.66
Re-tested, unseen data, median (%)12.92
Re-tested, unseen data, range (%)11.24 to 13.48
Excess caught, line and band (%)none
Excess caught, line alone (%)none