Haematology and immunology
CD4 absolute count, cells/µL
Flow cytometry, laboratory platform compared with Point of care CD4 and haemoglobin analyser. 247 paired results.
Compare it with the other comparisons in the explorer
Comparison line
The new system reads 3.7% higher than the old system on average. The comparison line (Passing-Bablok) has a slope of 1.044 (95% interval 1.014 to 1.071) and an intercept of -0.03.
Fig. 1. The comparison line against the line of no change, over the range of the decision limits (100, 200, 350, 500 cells/µL). Dots mark where each line meets each limit; the gap between them is the bias at that limit.
Crossing each limit
Results that fall on the other side of a decision limit on the new system than on the old, read as if the new system had replaced the old, against the number repeat testing on one unchanged system would move. The program reports the count across all limits together.
All decision limits together (100, 200, 350, 500 cells/µL)
Crossed between the two systems: 32
Expected from repeat testing alone: about 33
Grade and warnings
- Re-test band: full (at least the protocol minimum of pairs).
- Band half-width around each limit: 6.49%.
- Results re-tested on data the band was not fitted to, median: 15%.
- Repeat-test imprecision: default value.
- Note: Fewer than 5 results crossed beyond what repeat testing alone would move; excess caught is not meaningful here.
Source
Sagnia B, Mbakop Ghomsi F, Moudourou S, Gutierez A, Tchadji J, Sosso SM, Ndjolo A, Colizzi V (2024). Accurate and reproducible enumeration of CD4 T cell counts and Hemoglobin levels using a point of care system: Comparison with conventional laboratory based testing systems in a clinical reference laboratory in Cameroon. PLoS One. doi:10.1371/journal.pone.0297790 (opens in a new tab); PMC10954178 (opens in a new tab). Licence: CC BY 4.0 (opens in a new tab). Re-analysed by NakedSignal. The deposit names the systems.
Show the numbers
| Field | Value |
|---|---|
| Paired results used | 247 |
| Comparison line slope | 1.043741 |
| Slope, 95% interval | 1.014465 to 1.071434 |
| Comparison line intercept | -0.030269 |
| Mean difference, new minus old, as % of old | 3.734 |
| Decision limits (cells/µL) | 100, 200, 350, 500 |
| Crossed between the two systems | 32 |
| Expected from repeat testing alone | 33.04 |
| Band half-width (%) | 6.49 |
| Re-tested, unseen data, median (%) | 15.18 |
| Re-tested, unseen data, range (%) | 12.96 to 17 |
| Excess caught, line and band (%) | none |
| Excess caught, line alone (%) | none |